Influence of microstructure on the performance of nitinol: a computational analysis
View/ Open
Full Text
Date
2014-04-26Author
Weafer, F. M.
Bruzzi, M. S.
Metadata
Show full item recordUsage
This item's downloads: 0 (view details)
Cited 7 times in Scopus (view citations)
Recommended Citation
Weafer, F. M. Bruzzi, M. S. (2014). Influence of microstructure on the performance of nitinol: a computational analysis. Journal of Materials Engineering and Performance 23 (7), 2539-2544
Published Version
Abstract
One material that has found particular favor for use in biomedical endovascular stents is the near equi-atomic NiTi alloy, Nitinol. One remarkable trait exhibited by this superelastic material is the improvement of its fatigue performance with increasing mean strain (Ref 1, 2). Clarification into this phenomenon still remains incomplete in the literature. This study proposes a microstructural explanation for this unique macroscopic behavior; it is hypothesized stress-induced martensite (SIM) will stabilize with increasing strain which, in turn, leads to the observed increase in fatigue life. Finite element analysis (FEA) is employed to investigate the behavior of a "v-strut" stent subcomponent under various strain levels. The volume fraction of SIM is analyzed to identify its potential influence on macroscopic response and, ultimately, fatigue behavior. In addition, a computational investigation is performed on the effect of crystallographic texture on macroscopic response. Granular transformational behavior is analyzed using FEA models with realistic and idealized grain structures, specifically evaluating the effect of individual grain orientations on the stress-induced martensite transformation.