Show simple item record

dc.contributor.authorO'Connell, Sandra
dc.contributor.authorÓ Laighin, Gearóid
dc.contributor.authorKelly, Lisa
dc.contributor.authorMurphy, Elaine
dc.contributor.authorBeirne, Sorcha
dc.contributor.authorBurke, Niall
dc.contributor.authorKilgannon, Orlaith
dc.contributor.authorQuinlan, Leo R.
dc.date.accessioned2017-03-31T11:01:24Z
dc.date.available2017-03-31T11:01:24Z
dc.date.issued2016-05-11
dc.identifier.citationO’Connell S, ÓLaighin G, Kelly L, Murphy E, Beirne S, Burke N, et al. (2016) These Shoes Are Made for Walking: Sensitivity Performance Evaluation of Commercial Activity Monitors under the Expected Conditions and Circumstances Required to Achieve the International Daily Step Goal of 10,000 Steps. PLoS ONE 11(5): e0154956. doi:10.1371/journal.pone.0154956en_IE
dc.identifier.issn1932-6203
dc.identifier.urihttp://hdl.handle.net/10379/6417
dc.description.abstractIntroductionPhysical activity is a vitally important part of a healthy lifestyle, and is of major benefit to both physical and mental health. A daily step count of 10,000 steps is recommended globally to achieve an appropriate level of physical activity. Accurate quantification of physical activity during conditions reflecting those needed to achieve the recommended daily step count of 10,000 steps is essential. As such, we aimed to assess four commercial activity monitors for their sensitivity/accuracy in a prescribed walking route that reflects a range of surfaces that would typically be used to achieve the recommended daily step count, in two types of footwear expected to be used throughout the day when aiming to achieve the recommended daily step count, and in a timeframe required to do so.MethodsFour commercial activity monitors were worn simultaneously by participants ( n = 15) during a prescribed walking route reflective of surfaces typically encountered while achieving the daily recommended 10,000 steps. Activity monitors tested were the Garmin Vivofit (TM), New Lifestyles' NL-2000 (TM) pedometer, Withings Smart Activity Monitor Tracker ( Pulse O-2)(TM), and Fitbit One (TM).ResultsAll activity monitors tested were accurate in their step detection over the variety of different surfaces tested (natural lawn grass, gravel, ceramic tile, tarmacadam/asphalt, linoleum), when wearing both running shoes and hard-soled dress shoes.ConclusionAll activity monitors tested were accurate in their step detection sensitivity and are valid monitors for physical activity quantification over the variety of different surfaces tested, when wearing both running shoes and hard-soled dress shoes, and over a timeframe necessary for accumulating the recommended daily step count of 10,000 steps. However, it is important to consider the accuracy of activity monitors, particularly when physical activity in the form of stepping activities is prescribed as an intervention in the treatment or prevention of a disease state.en_IE
dc.description.sponsorshipSandra O’Connell is supported by the Irish Research Council grant number EPSPG/2012/504, under the Enterprise Partnership Scheme. The enterprise partner was the Galway Clinic, Doughiska, Co. Galway, Ireland.
dc.formatapplication/pdfen_IE
dc.language.isoenen_IE
dc.publisherPublic Library of Scienceen_IE
dc.relation.ispartofPlos Oneen
dc.subjectFree living conditionsen_IE
dc.subjectDwellingen_IE
dc.subjectOlder adultsen_IE
dc.subjectPhysical activityen_IE
dc.subjectPedometer accuracyen_IE
dc.subjectTreadmill walkingen_IE
dc.subjectHealthy adultsen_IE
dc.subjectValidityen_IE
dc.subjectReliabilityen_IE
dc.subjectDeviceen_IE
dc.titleThese shoes are made for walking: sensitivity performance evaluation of commercial activity monitors under the expected conditions and circumstances required to achieve the international daily step goal of 10,000 stepsen_IE
dc.typeArticleen_IE
dc.date.updated2017-03-27T13:28:08Z
dc.identifier.doi10.1371/journal.pone.0154956
dc.local.publishedsourcehttp://dx.doi.org/10.1371/journal.pone.0154956en_IE
dc.description.peer-reviewedpeer-reviewed
dc.contributor.funder|~|
dc.internal.rssid11140273
dc.local.contactGearóid Ó Laighin, Electronic Engineering, Room 3049, Engineering Building, Nui Galway. 2685 Email: gearoid.olaighin@nuigalway.ie
dc.local.copyrightcheckedNo
dc.local.versionACCEPTED
nui.item.downloads206


Files in this item

Attribution-NonCommercial-NoDerivs 3.0 Ireland
This item is available under the Attribution-NonCommercial-NoDerivs 3.0 Ireland. No item may be reproduced for commercial purposes. Please refer to the publisher's URL where this is made available, or to notes contained in the item itself. Other terms may apply.

The following license files are associated with this item:

Thumbnail

This item appears in the following Collection(s)

Show simple item record