Fluid dynamics of gas exchange in high-frequency oscillatory ventilation: In vitro investigations in idealized and anatomically realistic airway bifurcation models

View/ Open
Date
2008-11Author
Heraty, Kevin B.
Laffey, John G.
Quinlan, Nathan J.
Metadata
Show full item recordUsage
This item's downloads: 419 (view details)
Cited 20 times in Scopus (view citations)
Recommended Citation
Heraty, KB,Laffey, JG,Quinlan, NJ (2008) 'Fluid dynamics of gas exchange in high-frequency oscillatory ventilation: In vitro investigations in idealized and anatomically realistic airway bifurcation models'. Annals Of Biomedical Engineering, 36 :1856-1869.
Published Version
Abstract
The objective of this work is to develop understanding of the local fluid dynamic mechanisms that underpin gas exchange in high-frequency oscillatory ventilation (HFOV). The flow field during HFOV was investigated experimentally using particle image velocimetry in idealized and realistic models of a single bifurcation. Results show that inspiratory and expiratory fluid streams coexist in the airway at flow reversal, and mixing between them is enhanced by secondary flow and by vortices associated with shear layers. Unsteady flow separation and recirculation occurs in both geometries. The magnitude of secondary flow is greater in the realistic model than in the idealized model, and the structure of secondary flow is quite different. However, other flow structures are qualitatively similar.