dc.description.abstractThis paper discusses the impact of tissue dehydration on the dielectric properties of excised tissue samples. The effect of dehydration on the tissue surface has been characterized as a function of time after excision on freshly excised mouse liver. The dielectric properties of liver were measured over the frequency range of 500 MHz-20 GHz using an open-ended coaxial probe. Tissue samples were obtained from 7 athymic BALB/c Nude mice, and measurements were performed over the first 3.5 h post-excision at the tissue surface and in the middle of the sample (accessed via a small incision). The samples were kept in sealed containers between measurements to avoid excessive dehydration. The measured dielectric data show a change of more than 25% in both the real and imaginary parts of complex permittivity over 3.5 h after excision. Results indicate the impact of tissue dehydration on the dielectric properties, and signify the importance of considering controls in the experimental design of ex vivo dielectric measurements.
dc.publisherIOP Publishing
dc.relation.ispartofBiomedical Physics & Engineering Express
dc.rightsAttribution-NonCommercial-NoDerivs 3.0 Ireland
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/3.0/ie/
dc.subjectdielectric spectroscopy
dc.subjectelectromagnetic interactions
dc.subjectbiological tissue
dc.subjectbioimpedance
dc.subjecttissue dehydration
dc.subjectmicrowave-frequencies
dc.subjectelectrical-properties
dc.subjectbreast tissues
dc.subjectanimal-tissues
dc.subjectbrain-tissue
dc.subjectliver-tissue
dc.subjectin-vivo
dc.subject20 ghz
dc.subjectmhz
dc.subjectsystems
dc.titleInvestigation of the effect of dehydration on tissue dielectric properties in ex vivo measurements
dc.typeArticle
dc.identifier.doi10.1088/2057-1976/aa74c4
dc.local.publishedsourcehttp://iopscience.iop.org/article/10.1088/2057-1976/aa74c4/pdf
nui.item.downloads0


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record

Attribution-NonCommercial-NoDerivs 3.0 Ireland
Except where otherwise noted, this item's license is described as Attribution-NonCommercial-NoDerivs 3.0 Ireland