<title>Experimental and numerical assessment of MRI-induced temperature change and SAR distributions in phantoms</title>
<link href="http://hdl.handle.net/10379/6943" rel="alternate"/>
<author>
<name>Colgan, Niall</name>
</author>
<author>
<name>Van der Putten, Wil</name>
</author>
<author>
<name>Tuohy, Brendan</name>
</author>
<id>http://hdl.handle.net/10379/6943</id>
<updated>2017-10-28T01:01:09Z</updated>
<published>2017-06-01T00:00:00Z</published>
<summary type="text">Experimental and numerical assessment of MRI-induced temperature change and SAR distributions in phantoms
Colgan, Niall; Van der Putten, Wil; Tuohy, Brendan
Eskola, H., Väisänen, O., Viik, J., Hyttinen, J.
During an MR procedure, most of the transmitted RF power is transformed into heat within the patient’s tissue and implants&#13;
as a result of resistive losses which is referred to as the specific energy absorption rate (SAR) (1). The European committee&#13;
for electrotechnical standardisation (CENELEC) has mandated that all scanners must measure the specific absorption rate of&#13;
radiofrequency in patients and develop system safeguards to ensure that the limits set out IEC 60602-3-33 are not exceeded.&#13;
Accurate estimation of SAR is critical to safeguard in unconscious/sedated patients, patients with compromised thermoregulation,&#13;
implant patients, pregnant patients and neonates who require an MRI procedure. The increased static field strength and&#13;
RF duty cycle capabilities in modern MRI scanners means that systems can easily exceed safe SAR levels for patients (2).&#13;
Advisory protocols routine used to establish QA protocols do not have advise on the testing of SAR levels in MRI and this is&#13;
not routinely measured in annual medical physics QA. There is increasing need to verify the manufacturers SAR estimations.